SUP-825-J Signal Calibrator 0.075% High Accuracy

Intelligent industrial instrument calibrator. The signal calibrator is mainly used for measurement calibration and signal generation, with six working modes of measurement, generation, calibration, recording, communication and documentation, usually for m


Introduction

The 0.075% precision signal calibrator has a variety of signal output and measurement functions, including voltage, current and thermocouple, with LCD screen and silicone keyboard, simple operation, long standby time, high precision and programmable output. It is widely used in the debugging of LAB industry, PLC process instruments, electrical values and other fields.

Specification

General Specifications

Operating temperature

-10℃~55℃

Storage temperature

-20℃~70℃

Relative Humidity (%RH operating without condensation)

90%(10℃~30℃)

75%(30℃~40℃)

45%(40℃~50℃)

35%(50℃~55℃)

Uncontrolled<10℃

EMC

EN55022, EN55024

Vibration

Random, 2g, 5 to 500Hz

Concussion

30g, 11ms, half sine bow wave

Power requirement

4 AA Ni-MH, Ni-Cd batteries

Size

215mm×109mm×44.5mm

Weight

About 500g

 DC Voltage

 Range

Accuracy

 Measurement

 (0~100)mVDC(Upper display)

 ±0.02%

 (0~30)VDC(Upper display)

 ±0.02%

 (0~100)mVDC(Lower display)

 ±0.02%

 (0~20)VDC(Lower display)

 ±0.02%

 Source

 (0~100)mVDC

 ±0.02%

 (0~10)VDC

 ±0.02%

Resistance

Range

Accuracy

4-wire

2-, 3-wire

Accuracy

Accuracy

Measurement

(0~400)Ω

±0.1Ω

±0.15Ω

(0.4~1.5)kΩ

±0.5Ω

±1.0Ω

(1.5~3.2)kΩ

±1.0Ω

±1.5Ω

Excitation Current: 0.5mA

Clear of resistance before measuring according to  "10.4 Clear of Resistance and RTDs".

*3-wire: Assumes matched leads with a total resistance not exceeding 100Ω.

Resolution (0~1000)Ω: 0.01Ω;